This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Coverstainer Slide Rack
- Starting date:
- November 26, 2012
- Posting date:
- January 7, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-16625
Recalled Products
A. Coverstainer Slide Rack
Reason
The expected life time of the Coverstainer slide racks does not meet established performance requirements, originating from variations into the manufacturing process. The result can be that slides are falling out of the rack during the staining process.
Affected products
A. Coverstainer Slide Rack
Lot or serial number
All Lots
Model or catalog number
CS100
Companies
- Manufacturer
-
Dako Denmark A/S