Health product recall

Coulter DxH Diluent

Last updated

Summary

Product
Coulter DxH Diluent
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products

Lot or serial number

Model or catalogue number

Coulter DxH Diluent

More than 10 numbers, contact manufacturer.

628017

Issue

Beckman Coulter has confirmed that specific diluent lot numbers of Coulter DxH Diluent contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. This may result in elevated PLT background counts in Daily Checks ranging from 3.0 to 9.1 x 103 cells/μL.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies

Beckman Coulter, Inc.

250 S. Kraemer Blvd., Brea, California, United States, 92821

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81285