This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
COSEAL SURGICAL SEALANT (2016-05-10)
- Starting date:
- May 10, 2016
- Posting date:
- May 24, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-58526
Affected products
- COSEAL SURGICAL SEALANT
Reason
Baxter is issuing a recall for the product codes and lots listed, due to the potential for incomplete dissolution of the polyethylene glycol (PEG) component during the reconstitution of the product, which may affect the consistency of the hydrogel formation during use.
Affected products
A. COSEAL SURGICAL SEALANT
Lot or serial number
HA151205
HA160114
Model or catalog number
934071
934072
Companies
- Manufacturer
-
Baxter Healthcare Sa
Zurich
SWITZERLAND