Cortisol Saliva Luminescence Immunoassay (2018-01-12)
- Starting date:
- January 12, 2018
- Posting date:
- February 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65870
Affected products
Cortisol Saliva Luminescence Immunoassay
Reason
IBL detected an issue that indicates degradation in performance of the kit that results in elevated cortisol values. When testing patient samples with one of the affected lots, the controls and calibrator values will meet specifications.
Affected products
Cortisol Saliva Luminescence Immunoassay
Lot or serial number
- LCL 130
- LCL 134
Model or catalog number
- RE62111
- RE62119
Companies
- Manufacturer
-
IBL International GmbH
Flughafenstrasse 52A
Hamburg
22335
GERMANY