CorNeat KPro
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
CorNeat KPro |
Not applicable. |
ITA 359439 |
Issue
The recall is being carried out because cumulative global clinical data from the ongoing clinical investigation revealed emerging safety concerns. These concerns include:
- progressive conjunctival retraction, which in some cases led to partial exposure of the device.
- difficulties maintaining a complete ocular seal, compromising the intended function of the implant.
- a potentially increased risk of intraocular infection, associated with these complications.
Additional information
Details
Corneat Vision Ltd.
4 Hasheizaf St., Raanana, Israel, 4366411
Get notified