Coolief Sinergy Cooled Radiofrequency Kit (2016-06-17)

Starting date:
June 17, 2016
Posting date:
July 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59244

Affected products

A. Coolief Sinergy Cooled Radiofrequency Kit

Reason

A mismatch was identified between the length of the RF electrode (probe) and the cannula (introducer). There were 75 mm long introducers packaged with 100 mm probes, which if present and used, could potentially lead to an injury.

Affected products

A. Coolief Sinergy Cooled Radiofrequency Kit

Lot or serial number

M5264D205

Model or catalog number

SIK-17-75-4

Companies
Manufacturer

HALYARD HEALTH, INC.

5405 WINDWARD PARKWAY

ALPHARETTA

30004

Georgia

UNITED STATES