This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Cook Percutaneous Enteral Feeding Catheters (2017-10-19)

Starting date:
October 19, 2017
Posting date:
October 30, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-64996

Affected products

  1. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET
  2. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET MALLINCKRODT

Reason

Cook has identified that the affected lots may contain wire guides that were incorrectly loaded into the wire guide holder. This could lead to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Affected products

A. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET

Lot or serial number
  • G05274
  • G07637
Model or catalog number
  • WOGS-1200-WH-050493
Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES



B. WILLS-OGLESBY PERCUTANEOUS GASTROSTOMY SET MALLINCKRODT

Lot or serial number
  • G08417
Model or catalog number
  • WOGS-1400-MKDT-A
Companies
Manufacturer

Cook, Inc.

750 Daniels Way

Bloomington

47404

Indiana

UNITED STATES