This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Continuum MR Infusion System - Pump (2014-03-10)

Starting date:
March 10, 2014
Posting date:
March 25, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type I
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38611

Recalled Products

Continuum MR Infusion System - Pump

Reason

Incidents have been resulting in magnetic attraction caused by a damaged locking pin. The locking pin is found in the pump retaining bracket and is used to securely hold the pump in the bracket in the MR environment.

Affected products

Continuum MR Infusion System - Pump

Lot or serial number

3008236

3009135

3009917

3009966

Model or catalog number

More than 10 numbers, contact manufacturer 

Companies
Manufacturer

Medrad Inc.

625 Alpha Drive

Pittsburgh

15238

Pennsylvania

UNITED STATES