Health product recall

Concerto & AMP; Basic Shower Trolley (2019-08-08)

Starting date:
August 8, 2019
Posting date:
August 23, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70785


Last updated:

Affected Products

Concerto & AMP; Basic Shower Trolley

Reason

Arjo has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall, leading to possible injuries to patient and/or caregiver.

Affected products

Concerto & AMP; Basic Shower Trolley

Lot or serial number

Lots between P0512263  and P0551621.

Model or catalog number

BAB 1103-01

BAB1000-01

BAB5000-01

S8533570-031

Companies
Manufacturer

ArjoHuntleigh AB

Hans Michelsensgatan 10

Malmo

21120

SWEDEN