This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

COMPREHENSIVE INSTRUMENTATION 2 PRONG RETRACTOR (2016-02-19)

Starting date:
February 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57404

Affected Products

COMPREHENSIVE INSTRUMENTATION 2 PRONG RETRACTOR

Reason

Action following an investigation which identified that PN: 402582 was manufactured using 420 stainless instead of the material specified on the print, 420 S29 stainless steel. 420 stainless is more brittle than 420 S29. The instrument may fracture during use.

Affected products

COMPREHENSIVE INSTRUMENTATION 2 PRONG RETRACTOR

Lot or serial number

More than 20 numbers, contact manufacturer.

Model or catalog number

402852

Companies
Manufacturer
Biomet Orthopedics
56 East Bell Drive
Warsaw
46581
Indiana
UNITED STATES