CombiStar, SafeStar and TwinStar Filters/ HMEs
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Filter/HME TwinStar Plus |
All lots. |
MP05801 |
Filter SafeStar 90 Plus |
All lots. |
MP05785 |
Filter SafeStar 55 Plus |
All lots. |
MP05790 |
CombiStar Plus Filter + HME Hepa Flex |
All lots. |
MP12062 |
Filter SafeStar 60A Plus |
All lots. |
MP05795 |
Issue
Users have reported a slow increase of the CO2-curve when the gas sample line of the gas measurement of a Dräger anesthesia machine is connected to the gas sampling port of the affected filters/HME. The comparatively slower increase in the CO2-curve only occurs when the gas sample line is connected to the gas sampling port of an affected filter/HME. Other setups are not affected.
Recall start date: May 23, 2025