Health product recall

CombiDiagnost R90

Last updated

Summary

Product
CombiDiagnost R90
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

CombiDiagnost R90

All lots.

709030

Issue

If the CombiDiagnost R90 System has a damaged wire in the System Power Distribution Unit (SPDU) within the m-cabinet and is connected to the ground, or the system experiences a power surge, there is the potential for a short circuit of the Electromagnetic Interference (EMI) filter or UA/UB board. If this issue occurs, the user may experience one or more of the following:

- electrical fire within the main-cabinet (m-cabinet)

- see or smell smoke

- hear a popping sound coming from the m-cabinet

Recall start date: October 28, 2021

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems DMC GmbH

Roentgenstrasse 24, Hamburg, Germany, 22335

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74402

Get notified

Receive notifications for new and updated recalls and alerts by category.