This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Combi Dockable Table Neurosurgery

Starting date:
October 14, 2016
Posting date:
November 9, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60902

Affected Products

Reason

Siemens has reasons to believe that in some cases the Combi Dockable Table Neurosurgery may have a manufacturing defect. In certain cases a potential risk of a sudden drop in table height may exist. This can happen with or without user interaction and may occur unexpectedly. The maximum drop in height is 5 cm. To address this issue Siemens will inspect and correct any affected devices on site.

Affected products

Combi Dockable Table Neurosurgery

Lot or serial number

12811004

Model or catalog number

10684337

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY