Colleague & Spectrum Infusion Pumps (2021-03-17)
- Starting date:
- March 17, 2021
- Posting date:
- April 1, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75291
Last updated:
2021-04-01
Affected Products
- COLLEAGUE GUARDIAN ENHANCED CXE COLOR PUMP
- COLLEAGUE GUARDIAN ENHANCED 3 CXE COLOR PUMP
- FRENCH COLLEAGUE CXE VOLUMETRIC PUMP
- FRENCH COLLEAGUE 3 CXE VOLUMETRIC PUMP
Reason
Baxter is communicating important safety information regarding the practice of using multiple extension sets with infusion pumps. Increasing the length of IV tubing between the pump and the patient will result in varying increases in outlet pressure and/or decreases in intake pressure, which could lead to unknown and undetectable reductions in forward flow and flow rate accuracy.
Affected products
A. COLLEAGUE GUARDIAN ENHANCED CXE COLOR PUMP
Lot or serial number
- All serial numbers.
Model or catalog number
- 2M91617A
Companies
- Manufacturer
-
Baxter Healthcare SA
Zurich
8010
SWITZERLAND
B. COLLEAGUE GUARDIAN ENHANCED 3 CXE COLOR PUMP
Lot or serial number
- All serial numbers.
Model or catalog number
- 2M91637A
Companies
- Manufacturer
-
Baxter Healthcare SA
Zurich
8010
SWITZERLAND
C. FRENCH COLLEAGUE CXE VOLUMETRIC PUMP
Lot or serial number
- All serial numbers.
Model or catalog number
- DNM91617A
Companies
- Manufacturer
-
Baxter Healthcare SA
Zurich
8010
SWITZERLAND
D. FRENCH COLLEAGUE 3 CXE VOLUMETRIC PUMP
Lot or serial number
- All serial numbers.
Model or catalog number
- DNM91637A
Companies
- Manufacturer
-
Baxter Healthcare SA
Zurich
8010
SWITZERLAND