Health product recall

Colleague & Spectrum Infusion Pumps (2021-03-17)

Starting date:
March 17, 2021
Posting date:
April 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-75291



Last updated:
2021-04-01

Affected Products

  1. COLLEAGUE GUARDIAN ENHANCED CXE COLOR PUMP
  2. COLLEAGUE GUARDIAN ENHANCED 3 CXE COLOR PUMP
  3. FRENCH COLLEAGUE CXE VOLUMETRIC PUMP
  4. FRENCH COLLEAGUE 3 CXE VOLUMETRIC PUMP

Reason

Baxter is communicating important safety information regarding the practice of using multiple extension sets with infusion pumps. Increasing the length of IV tubing between the pump and the patient will result in varying increases in outlet pressure and/or decreases in intake pressure, which could lead to unknown and undetectable reductions in forward flow and flow rate accuracy.

Affected products

 

A. COLLEAGUE GUARDIAN ENHANCED CXE COLOR PUMP

Lot or serial number
  • All serial numbers. 
Model or catalog number
  • 2M91617A
Companies
Manufacturer

Baxter Healthcare SA

Zurich

8010

SWITZERLAND


B. COLLEAGUE GUARDIAN ENHANCED 3 CXE COLOR PUMP

Lot or serial number
  • All serial numbers. 
Model or catalog number
  • 2M91637A
Companies
Manufacturer

Baxter Healthcare SA

Zurich

8010

SWITZERLAND


C. FRENCH COLLEAGUE CXE VOLUMETRIC PUMP

Lot or serial number
  • All serial numbers. 
Model or catalog number
  • DNM91617A
Companies
Manufacturer

Baxter Healthcare SA

Zurich

8010

SWITZERLAND


D. FRENCH COLLEAGUE 3 CXE VOLUMETRIC PUMP

Lot or serial number
  • All serial numbers. 
Model or catalog number
  • DNM91637A
Companies
Manufacturer

Baxter Healthcare SA

Zurich

8010

SWITZERLAND