Codman Bactiseal EVD Catheter Set
Last updated
Summary
Product
Codman Bactiseal EVD Catheter Set
Issue
Medical devices - Sterility
What to do
Contact manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Codman Bactiseal EVD Catheter Set | 7322926 7322939 7322933 7314916 7322927 7322941 | 82-1745 82-1750 82-1749 |
Issue
The decision to conduct a voluntary removal of the product was based on the following: through an internal investigation, it was identified that there is a defect in the external sterile packaging (pouch) leading to a gap or/and weakness of the sealing. There is no issue with the inner sterile packaging (tray). However, even if the inner packaging (tray) is sealed correctly, the defect in the outer sterile packaging (pouch) could lead to cross contamination during handling.
Recall start date: May 16, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Integra Lifesciences Production Corporation |
11 Cabot Blvd., Mansfield, Massachusetts, United States, 02048 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75577
Get notified
Receive notifications for new and updated recalls and alerts by category.