Health product recall

Cocoon Convective Warming Machine (2019-11-18)

Starting date:
November 18, 2019
Posting date:
November 29, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71752


Last updated:

Affected Products

Cocoon Convective Warming Machine

Reason

Thermal damage occurred in some CWS4000 devices due to repeated insertion/removal and movement of power cord in the power entry module which lead to improper contact.  Fluid ingress may also have contributed to the issue.  Devices will be fitted with P-clips to immobilize power cord, and IFU updated.

Affected products

Cocoon Convective Warming Machine

Lot or serial number

All lots.

Model or catalog number

CWS 4000

CWS 5000

Companies
Manufacturer

Care Essentials PTY LTD.

25-27 Slevin Street, North Geelong

Victoria

AUSTRALIA