This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Cobas Liat Analyzer

Starting date:
August 9, 2017
Posting date:
September 5, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64426

Reason

The Cobas Liat System allows the use of several different barcode types for sample identification. If any of these barcode types are used together with Codabar or Code 39 barcodes and match an existing Codabar or Code 39 exactly, it is theoretically possible that a sample mismatch between the two samples could occur when using software version 3.0.

Affected products

Cobas Liat Analyzer

Lot or serial number

Not Applicable

Model or catalog number

07341920190

Companies

Manufacturer

Roche Molecular Systems Inc.

1080 US Highway 202

Branchburg

08876

UNITED STATES