COBAS IT MIDDLEWARE (2016-08-18)

Starting date:
August 18, 2016
Posting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60002

Affected Products

COBAS IT MIDDLEWARE

Reason

During the investigation of an issue from a customer complaint, it was found by the manufacturer that the daily cleanup sample can fail in deleting samples which are older than the configured retention period in the cobas IT middleware. Some samples which have been ordered from the lis and remain with open tests may not be deleted. Open tests are those for which results did not complete the processing cycle, including transmission to lis.

Affected products

COBAS IT MIDDLEWARE

Lot or serial number

All software versions.

Model or catalog number

06498027001
06498035001