This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

COBAS IT MIDDLEWARE (2016-08-18)

Starting date:
August 18, 2016
Posting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60002

Affected Products

COBAS IT MIDDLEWARE

Reason

During the investigation of an issue from a customer complaint, it was found by the manufacturer that the daily cleanup sample can fail in deleting samples which are older than the configured retention period in the cobas IT middleware. Some samples which have been ordered from the lis and remain with open tests may not be deleted. Open tests are those for which results did not complete the processing cycle, including transmission to lis.

Affected products

COBAS IT MIDDLEWARE

Lot or serial number

All software versions.

Model or catalog number

06498027001

06498035001

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY