Health product recall

Cobas H 232 Instruments (2019-10-22)



Starting date:
October 22, 2019
Posting date:
January 6, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71973


Last updated:

Affected Products

Cobas H 232 Instruments

Reason

Potential defect with the cardiac pipettes used with the cobas H 232 system. Please be advised that this urgent medical device correction (Roche ref. No. rr201929) is not a product removal, as the product itself can still be used as long as users follow the instruction provided in the warning stated in the action taken by Roche diagnostics.

In the past months Roche have received 9 customer complaints worldwide for cardiac pipettes (out of more than 11 million distributed single syringes). The following issues were observed: 8 cases where the needle came off the pipette and 1 case where the needle became stuck in a sample tube. The manufacturer has carried out an investigation and confirmed these findings. In all related complaints the issue was detected by the user and no injury occurred.

 

Affected products

Cobas H 232 Instruments

Lot or serial number

All lots.

Model or catalog number

4901142190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY