Health product recall

Cobas e801 Syringes (2018-03-23)

Starting date:
March 23, 2018
Posting date:
April 24, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66520

Affected products

Cobas e801 Syringes

Reason

The plunger of the Cobas e801 syringe assembly may be mounted in a tilted position which applies mechanical stress to the plunger and may cause r1, r2 or prewash liquids to be pipetted incorrectly.

Affected products

Cobas e801 Syringes

Lot or serial number
  • 1601-01 To 18E6-10
Model or catalog number
  • 07682913001
Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY