Health product recall

Cobas e411 Immunoassay Analyzer (2018-09-24)

Starting date:
September 24, 2018
Posting date:
October 19, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67948

Affected products

Cobas e411 Immunoassay Analyzer

Reason

Following investigation of customer complaints, the manufacturer changed the frequency interval for mandatory replacement of pinch valve tubing in order to avoid the risk to obtain potential too high results or out of the measuring range results.

Affected products

Cobas e411 Immunoassay Analyzer

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • 04775201001
  • 04775279001
Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY