Health product recall

Coaguchek XS Strips (2018-10-10)

Starting date:
October 10, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68214

Affected products

Coaguchek XS Strips

Reason

Since the new WHO reference standard rtf/16 is used for calibration of the CoaguChek XS strips, measurement deviations against non-Roche controls as well as laboratory methods for higher INR values have been documented. For values higher than 4.5 INR increasing positive bias was found for identified lots.

Affected products

Coaguchek XS Strips

Lot or serial number
  • 28124114
  • 28631922
  • 28632112
  • 28632311
  • 29415111
  • 29415122
  • 29494211
  • 29494511
  • 30497411
  • 31404312
  • 31404712
  • 32264115
Model or catalog number
  • 04625315019
  • 04625315172
  • 04625358019
  • 04625358172
  • 04625374190
Companies
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY