This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
CLOSED SUCTION SYSTEMS (2015-08-11)
- Starting date:
- August 11, 2015
- Posting date:
- August 24, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54790
Affected Products
A. KIMVENT NEONATAL / PEDIATRIC "Y" TRACHEAL SUCTION
B. KIMVENT MULTI-ACCESS PORT CLOSED SUCTION SYSTEM F
Reason
Halyard Health has recently received reports indicating that the thumb valve of a very small percentage of Kimvent Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "locked" position
Affected products
A. KIMVENT NEONATAL / PEDIATRIC "Y" TRACHEAL SUCTION
Lot or serial number
M5082T502
M5082T513
M5110T505
Model or catalog number
1910
196
198-5
Companies
- Manufacturer
-
Kimberly-Clark Global Sales LLC
1400, Holcomb Bridge Road
Roswell
ÉTATS-UNIS 30076
Georgia
UNITED STATES
B. KIMVENT MULTI-ACCESS PORT CLOSED SUCTION SYSTEM F
Lot or serial number
M5082T614
M5085T604
M5124T606
Model or catalog number
22103-5
22109-5
2260-5
Companies
- Manufacturer
-
Kimberly-Clark Global Sales LLC
1400, Holcomb Bridge Road
Roswell
ÉTATS-UNIS 30076
Georgia
UNITED STATES