Health product recall

Clip Gun System - Clip Gun (2017-12-18)

Starting date:
December 18, 2017
Posting date:
January 12, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65680

Affected products

Clip Gun System - Clip Gun

Reason

Testing performed by Medtronic showed that the outer sterile pouch of the product did not meet standards for package integrity. Specifically, the testing demonstrated that the outer pouch could be compromised and possibly be undetected by the user, which could potentially lead to patient infections. In an abundance of caution, these affected products are being removed from the field.

Affected products

Clip Gun System - Clip Gun

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • CG8900
  • CG8901
Companies
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES