This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CLIP APPLIER (2016-04-08)

Starting date:
April 8, 2016
Posting date:
April 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58174

Affected Products

CLIP APPLIER

Reason

From customer feedback, manufacturer was made aware of the inconsistent clip application. Although malformed clips are typically readily apparent to the user, this failure mode may lead to unoccluded vessels.

Affected products

CLIP APPLIER

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

CA090

Companies
Manufacturer

Applied Medical Resources Corporation

22872 Avenida Empresa

Rancho Santa Margarita

California

UNITED STATES