Health product recall

CLINITUBES with Capillary Caps & Safe CLINITUBES Blood Sampler (2020-03-03)



Starting date:
March 3, 2020
Posting date:
June 12, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73303


Last updated:

Affected Products

A. CLINITUBES with Capillary Caps

B. Safe CLINITUBES Blood Sampler

Reason

Radiometer Medical ApS has received an increased number of complaints stating that several customers have had troubles when using SafeClinitubes (942-892). The SafeClinitubes have little or no aspiration, which means that it is difficult or impossible to collect the patient blood with the SafeClinitubes. The lack of aspiration may lead to the hazardous situation of delayed availability of fetal scalp pH measurement results during the delivery process.

Affected products

A. CLINITUBES with Capillary Caps

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

100161

Companies
Manufacturer

Radiometer Medical Aps

Akandevej 21

Bronshoj

2700

DENMARK



B. Safe CLINITUBES Blood Sampler

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

400622

Companies
Manufacturer

Radiometer Medical Aps

Akandevej 21

Bronshoj

2700

DENMARK