Clinisys™ SQ Blood Bank
Brand(s)
Last updated
Summary
Product
Clinisys™ SQ Blood Bank
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Clinisys™ SQ Blood Bank (Sim Ovision F_Driver V 1.00 - 1.09) |
10.0.1 |
+B236SQBB202410/$$720241001P |
Issue
The issue occurs when a laboratory instrument sends an incomplete message for a sample resulting in a "timed out" error message for that initial sample. The driver can retain the partial data transfer in memory, leading to potential data merging with the subsequent sample's transmission.
Recall start date: April 14, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Clinisys, Inc.
3300 E Sunrise Drive, Tucson, Arizona, United States, 85718
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77797
Get notified
Receive emails about new and updated recall and safety alerts.