Health product recall

ClearSight Heart Reference Sensor (2017-12-11)

Starting date:
December 11, 2017
Posting date:
January 8, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65542

Affected Products

CLEARSIGHT HEART REFERENCE SENSOR

Reason

The current product as well as the affiliated labeling contains an expiration date of two years. Through investigation and analysis of complaint data, Edwards Lifesciences discovered that hrs units are susceptible to experiencing an "hrs out of range" error message in the EV1000 platform that are beyond six months from the date of manufacture.

Affected products

CLEARSIGHT HEART REFERENCE SENSOR

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

EVHRS
 

Companies
Manufacturer
Edwards Lifesciences LLC
1 Edwards Way
Irvine
92614
UNITED STATES