Health product recall

Citadel Plus (2018-08-10)

Starting date:
August 10, 2018
Posting date:
August 31, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-67612

Affected products

Citadel Plus

Reason

It has been identified that the power drive function of the citadel plus bed may potentially be operated with only the left-handle safety switch depressed on the condition that the bed is driven forward and encounters heavy resistance (i.e. steep slopes in excess of greater than 7 degrees). This is due to a software deficiency that has been detected after receiving several customer complaints.

Affected products

Citadel Plus

Lot or serial number

P0423627
P0431773

Model or catalog number

FXXXXXXXALVBB

Companies
Manufacturer
ArjoHuntleigh AB
Hans Michelsensgatan 10
Malmo
21120
SWEDEN