Health product recall

CIOS ALPHA (2016-05-24)

Starting date:
May 24, 2016
Posting date:
June 13, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58840

Affected products

A. CIOS ALPHA

Reason

During manufacturing process, it was discovered that the crimp-connection of a cable was faulty. In case the cable gets loose, the high voltage circuit of the system is discharging extremely slowly after powering off the system. This error may not lead to any harm to patients or users. The risk remains only to service personnel and only if they do not follow the described procedures for servicing the high voltage circuits in the device.

Affected products

A. CIOS ALPHA

Lot or serial number

10071

Model or catalog number

10308191

Companies
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY