This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

CIC-C1q - Circulating Immune Complexes (CIC) ELISA

Last updated

Summary

Product
CIC-C1q - Circulating Immune Complexes (CIC) ELISA
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

CIC-C1q - Circulating Immune Complexes (CIC) ELISA

3836

EK-CIC

Issue

In-house testing with CIC lot 3836 has revealed unexpectedly elevated values in internal control serum samples exceeding the expected target range. Notably, samples with target values below 5 μg eq/ml showed increased results of up to 50% above their target hence, values of samples around the recommended cut-off (5 μg eq/ml) obtained with this lot could incorrectly exceed the true value of the sample, and thus potentially lead to inaccurate conclusions.

Recall start date: January 17, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Immunology
Companies

Buhlmann Laboratories Ag

Baselstrasse 55, Schonenbuch, Switzerland, 4124

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76883

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe