Health product recall

Ci-Tro Heparin Controls Low and High (2018-12-18)

Starting date:
December 18, 2018
Posting date:
January 11, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68742

Affected products

Ci-Tro Heparin Controls Low and High

Reason

Siemens healthcare diagnostics has confirmed that the APTT ratio of the Dade Ci-Trol Heparin Control Low lot 548455 may recover below the expected range for the APTT ratio given within the instructions for use (ratio for Sysmex CA-1500 systems: 1.4 - 2.2). Internal investigation with this specific Dade Ci-Trol Heparin Control Low lot confirmed that the content of unfractionated heparin is very low. This may lead to a low recovery of the APTT ratio with most APTT reagents and also to non-detectable heparin neutralization after treatment with Dade Hepzyme. This renders this control lot unsuitable for some of the intended uses given in the instruction for use.

Affected products

Ci-Tro Heparin Controls Low and High

Lot or serial number

548455

Model or catalog number

B4224-50

Companies
Manufacturer
Siemens Healthcare Diagnostics Products Gmbh
Emil-Von-Behring-Str. 76
Marburg
35041
GERMANY