Chromophare Surgical Lighting System
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Chromophare Surgical Lighting System |
07613327296167 |
CH00000001 |
Issue
Stryker has discovered that certain lots of the soffit ring may not adequately support the weight of the ceiling cover. As a result, the cover may shift or slide down the ceiling tube. Movement of the ceiling cover can expose the area above the ceiling, which may allow particles, such as dust, debris, residue, or other contaminants, to enter the sterile field.