Health product recall

Centricity Universal Viewer Zero Footprint Client (2018-11-08)

Starting date:
November 8, 2018
Posting date:
November 30, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68482

Affected products

Centricity Universal Viewer Zero Footprint Client

Reason

Image orientation markers may be displayed incorrectly in certain scenarios when images are flipped within centricity universal viewer Zero Footprint (ZFP).
Issue #1 affects software versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.A AND SP8.0.2.
When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cine; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect.
Issue #2 affects software versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.A, SP8.0.2, SP9 AND SP9.0.1.
When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.

Affected products

Centricity Universal Viewer Zero Footprint Client

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

00840682102988

Companies
Manufacturer
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES