Centricity Universal Viewer Zero Footprint Client (2018-11-08)
- Starting date:
- November 8, 2018
- Posting date:
- November 30, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68482
Affected products
Centricity Universal Viewer Zero Footprint Client
Reason
Image orientation markers may be displayed incorrectly in certain scenarios when images are flipped within centricity universal viewer Zero Footprint (ZFP).
Issue #1 affects software versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.A AND SP8.0.2.
When an image is flipped horizontally and vertically, the orientation markers appear correctly on the image to which they were applied. However, once new image rendering happens on this image or series (with scrolling; playing cine; maximizing or changing viewports/layouts; or synchronizing viewports and resizing the browser), the new image orientation is correctly maintained but the display of orientation markers can become incorrect.
Issue #2 affects software versions 6.0 SP7, SP7.0.1, SP7.0.2, SP7.0.3, SP8, SP8.0.1, SP8.0.A, SP8.0.2, SP9 AND SP9.0.1.
When an image in a series is flipped and then rotated in Zero Footprint Viewer (ZFP), the orientation markers appear correctly on the image to which they were applied. However, once you navigate to another image in the series and the image you navigated to has a different orientation than the prior image from which you navigated, the new image orientation is correctly maintained but the display of orientation markers can become incorrect.
Affected products
Centricity Universal Viewer Zero Footprint Client
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
00840682102988
Companies
- Manufacturer
-
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES