Health product recall

Centricity Universal Viewer (2018-10-11)

Starting date:
October 11, 2018
Posting date:
November 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68162

Affected products

Centricity Universal Viewer

Reason

An error can occur in Centricity Universal Viewer during study modification (delete/merge/split) activities that prevents synchronization of these updates with the Centricity Enterprise Archive (EA) database. When a diagnostic viewer (e.g. Centricity Universal Viewer Zero Footprint) is connected directly to the Centricity EA and is used for image interpretation, there is the possibility of viewing studies with incorrect patient images because the archive does not contain the updated study.

Affected products

Centricity Universal Viewer

Lot or serial number

2090255-001

Model or catalog number

2088034-X

Companies
Manufacturer
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES