Centricity PACS Workstation Software (2014-03-31)
- Starting date:
- March 31, 2014
- Posting date:
- March 20, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-52571
Recalled products
- Centricity PACS Workstation Software
Reason
In a reject image workflow where multiple studies are simultaenously opened on the centricity PACS RA1000 workstation and images are rejected from the active displayed study, images from the previously displayed study may also be rejected. The issue can occur with a previously displayed study that remains open in the background.
Affected products
A. Centricity PACS Workstation Software
Lot or serial number
More than 10 numbers, please contact the manufacturer.
Model or catalog number
- K2000YW
Companies
- Manufacturer
-
GE Healthcare
540 W. Northwest Highway
Barrington
Illinois
UNITED STATES