This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Centricity PACS Workstation Software (2014-03-31)

Starting date:
March 31, 2014
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52571

Recalled products

  1. Centricity PACS Workstation Software

Reason

In a reject image workflow where multiple studies are simultaenously opened on the centricity PACS RA1000 workstation and images are rejected from the active displayed study, images from the previously displayed study may also be rejected. The issue can occur with a previously displayed study that remains open in the background.

Affected products

A. Centricity PACS Workstation Software

Lot or serial number

More than 10 numbers, please contact the manufacturer.

Model or catalog number
  • K2000YW
Companies
Manufacturer
GE Healthcare
540 W. Northwest Highway
Barrington
Illinois
UNITED STATES