Centricity PACS-IW (2018-02-23)
- Starting date:
- February 23, 2018
- Posting date:
- March 19, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-66182
Affected products
Centricity PACS-IW
Reason
A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more "image series" (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR thorax exam with 1 image per series. There have been no injuries reported as a result of this issue.
Affected products
Centricity PACS-IW
Lot or serial number
00000222448II3
Model or catalog number
2050578-X
Companies
- Manufacturer
-
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES