Health product recall

Centricity PACS-IW (2018-02-23)

Starting date:
February 23, 2018
Posting date:
March 19, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66182

Affected products

Centricity PACS-IW

Reason

A database handling error could occur during the image acquisition process affecting the completeness of acquired images with Centricity PACS-IW. There is a potential that one or more "image series" (i.e. all images within an image set) may be missing from an exam without indication to the user. While this is rare, this can occur with imaging studies that consist of a very small number of images per series. E.g. CR thorax exam with 1 image per series. There have been no injuries reported as a result of this issue.

Affected products

Centricity PACS-IW

Lot or serial number

00000222448II3

Model or catalog number

2050578-X

Companies
Manufacturer
GE Healthcare
500 W Monroe Street
Chicago
60661
Illinois
UNITED STATES