This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Centricity Laboratory - Core Lab (2013-09-03)

Starting date:
August 30, 2013
Posting date:
October 28, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36429

Recalled Products

Centricity Laboratory - Core Lab 

Reason

Software error where when the user set micro_correct_msg_type and correct_msg_type to different variables in microbiology result entry when released results are changed and the correct results mesages are edited on the free-text form, some of the originally entered panel free-text is lost.

Affected products

Centricity Laboratory - Core Lab 

Lot or serial number

Not applicable

Model or catalog number
  • CORE
Companies
Manufacturer

GE Healthcare Inc.

540 West Northwest Hwy.

Barrington

60010

Illinois

UNITED STATES