Centricity Laboratory - Core Lab (2013-09-03)

Starting date:
August 30, 2013
Posting date:
October 28, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36429

Recalled Products

Centricity Laboratory - Core Lab 

Reason

Software error where when the user set micro_correct_msg_type and correct_msg_type to different variables in microbiology result entry when released results are changed and the correct results mesages are edited on the free-text form, some of the originally entered panel free-text is lost.

Affected products

Centricity Laboratory - Core Lab 

Lot or serial number

Not applicable

Model or catalog number
  • CORE
Companies
Manufacturer
GE Healthcare Inc.
540 West Northwest Hwy.
Barrington
60010
Illinois
UNITED STATES