Centrella Bed Frame (2020-05-12)
- Starting date:
- May 12, 2020
- Posting date:
- May 29, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73213
Last updated: 2020-05-29
Affected Products
Centrella Bed Frame
Reason
This recall related to Centrella Smart+ bed with software version 1.27.000 when connected to a Rauland Responder 5 nurse call system. The bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.
Affected products
Centrella Bed Frame
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
Not applicable.
Companies
- Manufacturer
-
Hill-Rom Inc.