Health product recall

Central Water Plant (CWP) Reverse Osmosis 208V Units (2020-03-23)



Starting date:
March 23, 2020
Posting date:
April 17, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72807


Last updated:

Affected Products

Central Water Plant (CWP) Reverse Osmosis 208V Units

Reason

Baxter Corporation is issuing an urgent medical device correction for the Central Water Plant (CWP) Reverse Osmosis 208v units in order to correct devices where there is potential for insufficient crimping of the cables within the main pump. The issue was identified as a result of customer complaints where the top of the main pump had visible burn marks due to heat build-up.

 

Affected products

Central Water Plant (CWP) Reverse Osmosis 208V Units

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number
  • WRO 131 ROHH
  • WRO 132 ROHH
Companies
Manufacturer

Gambro Lundia AB

MAGISTRATSVAGEN 16, P.O. BOX 10101,

Lund

22010

SWEDEN