Health product recall

CellMek SPS

Last updated

Summary

Product
CellMek SPS
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

CellMek SPS

SW Version: 2.3.90.0

C44603

Issue

Beckman Coulter has confirmed an error in SW version 2.3.90.0 that causes samples to be dispensed unexpectedly close to the bottom of the output tubes. The intended design is for samples to be dispensed above the fluid surface into the output tubes.
 

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Haematology
Companies
Beckman Coulter, Inc.
250 S. Kraemer Blvd., Brea, California, United States, 92821
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81436

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe