Ceftazidime 256 M.I.C.E

Starting date:
August 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60354

Affected products

Reason

A technical investigation has confirmed that CT1629B CAZ10, Ceftazidime batch 1426486 may contain discs with a reduced concentration of antibiotic that may impact performance testing with Pseudomonas aeruginosa ATCC 27853 (European Committee on Antimicrobial Susceptibility Testing (EUCAST) Quality Performance).

Continued used of this lot could result in quality control failures, delayed results or incorrect results reporting (false resistance).

Affected products

Ceftazidime 256 M.I.C.E

Lot or serial number

1426486

Model or catalog number

MA0120D

MA0120F

Companies
Manufacturer

Oxoid Limited

Wade Road

Basingstoke, Hampshire

RG24 8PW