This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Cedia TDM Assay - Tobramycin II (2013-10-16)

Starting date:
October 16, 2013
Posting date:
April 22, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39157

Recalled Products

Cedia TDM Assay - Tobramycin II

Reason

The reconstituted reagents do not maintain 30 days reconstituted stability. The instability of the reconstituted reagents may cause higher recovery of samples.

Affected products

Cedia TDM Assay - Tobramycin II

Lot or serial number

60169764

Model or catalog number

100018

Companies
Manufacturer

Microgenics Corporation

46360 Fremont Blvd.

Fremont

94538

California

UNITED STATES