Carina Ventilators
Brand(s)
Last updated
Summary
Product
Carina Ventilators
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Carina Ventilators |
More than 10 numbers, contact manufacturer. |
5704110-XX |
Issue
The recall is being initiated as there is a potential contamination of breathing gas. During standard tests, concentrations of 1,3-dichloropropan-2-ol were measured that exceed the acceptable uptake level during continuous use (>30 days in paediatric patients).
Recall start date: June 23, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Draegerwerk AG & Company KGAA
Moislinger Allee 53-55, Lubeck, Germany, 23542
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73950