Carestation 650 (2018-07-24)
- Starting date:
- July 24, 2018
- Posting date:
- August 17, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-67384
Affected products
Carestation 650
Reason
On systems equipped with the Auxiliary Common GGAS Utlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia. It would still be possible to ventilate the patient using the circle breathing system in the manual bag or mechanical ventilation modes.
Affected products
Carestation 650
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
1012-9650-000
Companies
- Manufacturer
-
Datex-Ohmeda, Inc.
3030 Ohmeda Drive, P.O. Box 7550
Madison
53707-7550
Wisconsin
UNITED STATES