Health product recall

CareLink 2090 Programmer and CareLink Encore 29901 Programmer (2018-10-08)

Starting date:
October 8, 2018
Posting date:
November 2, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68172

Affected products

  1. CareLink 2090 Programmer
  2. CareLink Encore 29901 Programmer

Reason

Purpose of this recall is to notify customers regarding modification to improve cybersecurity of device programmers by changing how the programmers are updated with new software. Medtronic will be disabling the SDN for programmer updates and will rely solely on the USB update method. Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download. To date, Medtronic has received zero (0) reports to indicate that such an issue has occurred.

Affected products

A. CareLink 2090 Programmer

Lot or serial number

All lots.

Model or catalog number

2090

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



B. CareLink Encore 29901 Programmer

Lot or serial number

All lots.

Model or catalog number

29901A

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES