Care Plus Incubator
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Care Plus Incubator |
More than 10 numbers, contact manufacturer. |
3000 |
Issue
GE Healthcare has become aware that the bedside panel of the incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the
bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
Recall start date: February 5, 2024
Additional information
Details
Datex-Ohmeda, Inc. |
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226 |