Care Plus Incubator
Brand(s)
Last updated
Summary
Product
Care Plus Incubator
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Care Plus Incubator | More than 10 numbers, contact manufacturer. | 3000 4000 1000 |
Issue
GE Healthcare has become aware that the bedside panel of the incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the
bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
Recall start date: February 5, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Obstetrics and gynaecology
Companies
Datex-Ohmeda, Inc. |
9900 Innovation Drive, Wauwatosa, Wisconsin, United States, 53226 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75073