Health product recall

CardioMEMS™ Patient Electronics System

Brand(s)
Last updated

Summary

Product
CardioMEMS™ Patient Electronics System
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number/

Model or catalog number

CardioMEMS™ Patient Electronics System

Not applicable.

CM1100

Issue

Abbott recently observed an increase of reports (complaints) in which the power connector plug became frayed or damaged over time. Damage and fraying of the power connector plug can occur with repeated bending or manipulation beyond 90 degrees. Fraying can rarely result in wires becoming exposed. The current report (complaint) rate has remained very low.

Recall start date: October 4, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Abbott Medical

387 Technology Circle  Nw, Atlanta, Georgia, United States, 30313

Published by
Health Canada
Audience
General public
Healthcare
Recall class
Type II
Identification number
RA-74514

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe