Health product recall

Cardiomems Hospital Electronics System

Brand(s)
Last updated

Summary

Product
Cardiomems Hospital Electronics System
Issue
Medical devices - Performance issue
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Cardiomems Hospital Electronics System All lots. CM2000
Cardiomems Hospital Electronics System All lots. CM3000

Issue

The Cardiomems HF System is intended to operate at frequencies between 30 and 37.5 MHz. Abbott has found that a small number (less than 1%) of implanted Cardiomems PA sensors (model CM2000) have operated outside of this intended radiofrequency range at least once over the life of the implant when interrogated by Cardiomems patient or hospital systems (models CM1100 and CM3000). Operating frequency may be affected by atmospheric conditions, elevation, pulmonary artery (pa) pressure, and unique sensor and patient characteristics.

Recall start date: February 7, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Abbott Medical

387 Technology Circle  Nw, Atlanta, Georgia, United States, 30313

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72929

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe