Cardinal Health Level 3 Isolation Gowns
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Cardinal Health SMS Elastic Cuff Thumbhook Level 3 Isolation Gown, Full-Back |
All lots. |
AT6688-2XL |
Convertor Isolation Gown, AAMI Level 3 Protection |
All lots. |
AT6200 |
Issue
The listed product codes and all lots of Cardinal Health Level 3 Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (the Association for the Advancement of Medical Instrumentation) level 3 protection requirements. The raw material identified has demonstrated possible degradation due to exposure to lighting, temperature, and relative humidity beyond the manufacturer recommended limits. This product recall is specific to the codes listed. Other codes are not impacted.
Recall Start Date: November 5, 2024
Additional information
Details
Cardinal Health 200, Llc Also Trading As Cardinal Health
3651 Birchwood Drive, Waukegan, Illinois, United States, 60085
Get notified